Leave Your Message
Particulate matter sampler ZR-3924
Environmental monitoring products

Particulate matter sampler ZR-3924

It uses filter membrane to capture particles in the ambient air (TSP, PM10, PM2.5).

  • Ambient air sampling flow (15~130)L/min
  • Load capacity When the flow is 100L / min, the load capacity is > 6kpa
  • A/B/C/D channels Atmospheric sampling flow (0.1~1.5)L/min
  • Power Supply AC(220±22)V, (50±1)Hz
  • Size (L310×W150×H220)mm
  • Weight About 5.0kg (battery include)
  • Power consumption ≤120W

product detail

Overview

ZR-3924 Particulate matter sampler is a portable instrument. It uses filter membrane to capture particles in the ambient air (TSP, PM10, PM2.5). The solution absorption method is used to collect various harmful gases in the ambient atmosphere and indoor air. It can be used for aerosol monitoring by environmental protection, health, labor, safety supervision, scientific research, education and other departments.

Features

• 4.3-inch color screen, touch operation and the operation is simple
• Compact size, light in weight, easy to carry
• Built-in Lithium Battery
• Four channels of simultaneous sampling can be used to collect particulate matter and gaseous pollutants in the air
• Rain proof, dust-proof, anti-static and anti-collision design can ensure normal operation under the conditions of rain, snow, dust and heavy haze
• Various sampling methods, such as constant sampling time, continuous sampling time and 24-hour sampling, can be realized
The cutter(TSP / PM 10 / PM 2.5) is made of aluminum alloy with anti-static adsorption
• Power-off memory function, continue to sample procedure when recovery
• Support data storage and export data with USB
• Print with wireless Bluetooth

Accessories

peitu-01.png

peitu-02.png

Standards

ISO 14644

ISO 21501-4:2018

GMP

GB/T 16292-2010

JIS B 9921:2010 

GB/T 6167-2007

JJF 1190-2008

Features


Technical Parameter

Parameter

Range

Resolution

Error

Ambient air sampling flow

(15130)L/min

0.1L/min

±5.0%

Ambient air sampling time

1min99h59min

1s

±0.1%

Load capacity

When the flow is 100L / min, the load capacity is > 6kpa

A/B/C/D channels

Atmospheric sampling flow

(0.11.5)L/min

0.01L/min

±2.0%

Atmospheric sampling time

1min99h59min

1s

±0.1%

Ambient atmospheric pressure

(60130)kPa

0.01kPa

±0.5kPa

Temperature range of incubator

(15-30)

0.1

±2

Noise

65dB(A)

Discharge duration

Three circuits work at the same time, TSP load is 2KPa, and the discharge time is>8h

Charging time

Internal charging14h, external charging5h

Power Supply

AC(220±22)V, (50±1)Hz

Size

(L310×W150×H220)mm

Weight

About 5.0kg (battery include)

Power consumption

120W