Leave Your Message
Airborne Particle Counter 3.53CFM ZR-1640

Biosafety cabinet & clean room

Airborne Particle Counter 3.53CFM ZR-1640

ZR-1640 Airborne Particle Counter is a portable precision particle counter.Built in vacuum pump, the flow is stably controlled at 3.53CFM(100 L/min).

  • Particle Size 0.3、 0.5、1.0、3.0、5.0、10.0μm
  • Counting Efficiency 0.3μm:50%; >0.5μm:100%
  • Error of particle size 0.5μm、5μm size ≤±30%
  • Concentration Limits 3500000/m3、10% coincidence loss

The instrument uses light scattering principle to measure the particle size and quantity in the air which the particle size is 0.3μm~10.0 μm. It is mainly used in clean room testing, air filter and filtering material performance testing and other fields.

It can be mainly used in cleanroom / operating room monitoring and verification, filter testing, IAQ investigation, data center cleaning and other fields.

xiangqing2.jpg


ISO 14644-9:2022 Cleanrooms and associated controlled environments

ISO 21501-4:2023 Light Scattering Airborne Particle Counter for Clean Spaces

JIS B 9921:2010 Light scattering airborne particle counter for clean spaces

GMP

Six channels simultaneous sampling and detection

Zero count less than 5 minutes

Built in vacuum pump, the flow is stably controlled at 3.53CFM(100 L/min).

Provide isokinetic sampling probe, temperature and humidity sampling probe

Record accumulated count value of each channel automatically

Built in HEPA filter to filter the exhaust gas

Historic data can be exported by USB or printed by Bluetooth printer

Record ambient temperature humidity atmospheric pressure

> Built-in lithium-ion battery, continuous sampling for about 8 hours when fully charged

7-inch color screen, wide working temperature , clear Visual in sunshine.

Parameter

Range

Particle Size

0.3、 0.5、1.0、3.0、5.0、10.0μm

Counting Efficiency

0.3μm:50%;   >0.5μm:100%

Error of particle size

0.5μm、5μm size ≤±30%

Repeatability

≤10%FS at the same test condition

Concentration Limits

3500000/m3、10% coincidence loss

Laser Source

Laser diodes

Zero Count Level

<1 count/5min

Flow Rate

100 L/min,maximum permissible error ±2%

Count Mode

Automatic;Accumulate

Count Time

1~36000s

Count Cycle

1~100 times

Exhaust Filtration

Build-in HEPA filter(>99.97%@0.3μm)

Operation Environment

(-20~50)℃, ≤85%RH

Power Adapter

DC12V,2A

Battery

Rechargeable Lithium-ion battery 25.9V12Ah

Battery Run Time

About 4 hours

Battery Charge Time

About 5 hours

Noise

<60dB(A)

Host size

(310×275×230)mm

Weight

About 7.5kg

Power Consumption

About 80W