ZR-3930B Environmental Analysis Automatic Sampler
Ambient air particulate matter sampler,equipped with particle separate device in line with national and American EPA standards is used for PM10,PM2.5 date quality control comparison of air automatic station.
Standards
HJ 618-2011 Ambient air PM10 and PM2.5 gravimetric method
EPA Air Quality Criteria for Particulate Matter
Features
> Compliance with US EPA equivalent method.
> The control system is stable, so that the flow rate is stable at 16.7L/min.
> Diversified cutters, with optional impact or cyclone cutters; Automatic film change under unattended operation.
> Intelligent temperature control system to ensure sampler operation under extreme weather conditions.
> Waterproof and dust proof.
> The sampling group,time and period can be set flexibly.
> Support wireless (GPRS) and wired (RS-232/USB) data transmission, remote monitoring via PC and SMS notification with mobile phone.
> Automatic measurement and storage of various environmental parameters.
Deliver Goods
Parameter |
Range |
Resolution |
Accuracy |
||
Sampling flow |
16.7L/min |
0.01L/min |
±2.0% |
||
Sampling time |
1min~99h59min |
1s |
±0.2% |
||
Atmospheric pressure |
(60~130)kPa |
0.01kPa |
±0.5kPa |
||
Resistance |
Resistance > 35Kpa at 16.7L / min flow |
||||
Size of the filter membrane |
Φ47mm |
||||
PM10 cutter |
Da50=(10±0.5)μm, σg=1.5±0.1 |
||||
PM2.5 cutter |
Da50=(2.5±0.2)μm, σg=1.2±0.1 |
||||
Data storage |
1000 |
||||
Noise |
≤55dB(A) |
||||
Size |
20 membranes |
(L495×W280 ×H 680)mm |
10 membranes |
(L495×W280×H540)mm |
|
Weight |
About 25kg |
About 20kg |
|||
Power consumption |
≤200W |
≤100W |