ZR-1000C Mask Bacterial Filtration Efficiency (BFE) tester
Key specifications of Model ZR-1000 mask bacteria filtration efficiency tester (BFE) not only meet the requirements of B.1.1 in YY0469-2011 <technical requirements for medical surgical masks>, but also meet the requirements of ASTMF2100, ASTMF2101 and European EN14683 standards. What’s more, we made innovative improvements based on this, with double gas contrast sampling method at the same time, improve the precision of sampling. It is applicable to the performance test of the mask bacterial filtration efficiency by the metrological verification department, scientific research institutes, mask manufacturers and other relevant departments.
Features
> The negative pressure test system ensures the safety of operators;
> The negative pressure cabinet is equipped with peristaltic pump, A and B two-path six stage Anderson sampler;
> The flow of peristaltic pump can be set;
> The flow rate of special microbial aerosol generator can be set.
> Good nebulization performance, aerosol size conforms to global standards;
> Controlled by embedded high-speed industrial microcomputer;
> 10.4-inch industrial high brightness colored touch screen;
> USB port, supporting data transfer to USB flash disk;
> Support Bluetooth printing;
> Built-in high brightness lights;
> Built in leakage protection switch to protect the safety of operators;
> The inner layer of the cabinet is made of stainless steel, and the outer layer is cold-rolled sheet sprayed by plastic;
> Heat preservation and flame retardant is set between inner and outer layers;
> The front openable glass door is convenient for the experimental operator to observe and operate;
> Detachable holder, and the caster can be used for supporting and moving.
> Retractable rotating display screen, convenient to operate;
> The negative pressure cabinet has a built-in operating key board, which is corrosion-resistant and UV-resistant, make it convenient for operations.
Standards
EN14683-2019
ASTM F2100, STM F2101
YY 0469-2011
YY/T 0969-2013
GB/T 38880-2020
T/CNTAC 55—2020 T/CNITA 09104——2020
T/CTCA 7—2019
Deliver Goods
Main parameters |
Parameter range |
Resolution |
MPE (Maximum Permissible Error) |
Sample Flow Rate of A |
28.3L/min |
0.1L/min |
±2.0% |
Sample Flow Rate of B |
28.3L/min |
0.1L/min |
±2.0% |
Nebulization flow rate |
(0.1~10)L/min |
0.1L/min |
±5.0% |
Peristaltic pump flow |
(0.001~3.0) mL/min |
0.01mL/min |
±2.0% |
Front pressure of A-path flowmeter |
(-20~0)kPa |
0.1kPa |
±2.5% |
Front pressure of B-path flowmeter |
(-20~0)kPa |
0.01kPa |
±2.5% |
Front pressure of nebulizer |
(0~50)kPa |
0.01kPa |
±2.5% |
Negative pressure in aerosol chamber |
(-90~-120)Pa |
0.1Pa |
±1.0% |
Operation temperature |
(0~50)℃ |
||
Cabinet negative pressure |
(-50~-200)Pa |
||
Data storage capability |
>500000 groups |
||
Characteristics of HEPA filter |
≥ 99.995%@0.3μm, ≥ 99.995%@0.12μm |
||
Mean mass diameter of aerosol |
Average particle diameter is (3.0 ± 0.3)μm, geometric standard deviation ≤ 1.5 |
||
Captured particle size of Two-path 6-stage Anderson sampler |
1st stage > 7 μm, 2nd stage (4.7-7) μm,3rd stage (3.3-4.7) μm, 4th stage(2.1-3.3) μm, 5th stage (1.1-2.1) μm, 6th stage (0.6-1.1) μm |
||
Size of aerosol chamber |
(length 600× diameter 85× thickness 3) mm |
||
Total number of positive quality control sampler |
(2200±500) CFU/mL |
||
Ventilation flow rate of negative pressure cabinet |
≥5m3/min |
||
Dimensions of negative pressure cabinet door |
(length 1100× width 780) mm |
||
Host size |
(length 1050× width 635× height 1310) mm |
||
Bracket size |
(length 1050× width 635× height 650) mm |
||
Power supply |
AC220V±10%,50Hz |
||
Instrument noise |
<65dB(A) |
||
Weight of whole machine |
About 180 kg |
||
Overall power consumption |
<500W |