ZR-5410A is a Portable Comprehensive Calibrator for Gas, dust, smoke dust Sampler.

  • Soap film flowmeter (50~6000)mL/min
  • Roots flowmeter (6~260)L/min
  • Medium flow orifice flowmeter (40~160)L/min
  • High flow orifice flowmeter (700~1400)L/min
  • Size (L232×W334×H215)mm
  • Host Weight About 9kg

ZR-5410A is a Portable Comprehensive Calibrator for Gas, dust, smoke dust Sampler. Especially for calibrating the flow rate and pressure of air sampler, particulate matter sampler and flue gas analyzer.

Applications >

> Calibration service companies and service industry

> Measurement and control laboratories

> Quality assurance

1)High measurement accuracy

> The maximum error of flowrate is ±1% (First-Glass Standard) 


2)Meet the multi-type flow calibrator

> Built-in roots flowmeter to calibrate flue gas analyzer, can read flow directly.

> Built-in soap film flowmeter to calibrate air sampler and flue gas sampler.

> Built-in orifice flowmeter was used to calibrate the particulate sampler.


3)Excellent human interaction experience

> Built-in high-performance lithium battery, power supply time > 8h.

> Ambient atmospheric pressure, temperature, can be measured and input.

> Automatic conversion of standard flow.

> LCD screen, easy to operate.

Parameter

Range

Resolution

Accuracy

Soap film flowmeter

(50~6000)mL/min

0.1mL/min

±1.0%

Roots flowmeter

(6~260)L/min

0.01L/min

±1.0%

Medium flow orifice flowmeter

(40~160)L/min

0.01L/min

±1.0%

High flow orifice flowmeter

(700~1400)L/min

0.1L/min

±1.0%

Atmospheric temperature

(-20~50)℃

0.1℃

±1.0℃

Micro-pressure

(0~3000)Pa

1Pa

±1%

Gauge pressure

(-50~50)kPa

0.01kPa

±2%

Atmospheric pressure

(60~130)kPa

0.01kPa

±0.5kPa

Repeatability of flowrate test

±0.5%

Battery

>8h

Power supply

AC(100~240)V, 50/60Hz, DC12V 2A

Size

(L232×W334×H215)mm

Host Weight

About 9kg

Consumption