ZR-7250 Air Quality Monitoring Station
Air Quality Monitoring Station (AQMS) is a system that measures metrological parameters such as temperature, humidity, barometric pressure, wind speed, wind direction, noise and ambient parameters. The AQMS also integrates a series of ambient analyzers to monitor the concentration of air pollutants (such as SO2, NOX, CO, O3, PM10, PM2.5 etc.) real-time and continuously.
Suitable for use on a range of projects, including national and urban air monitoring networks, roadside monitoring, and industrial perimeter monitoring.
Who is ZR-7250 for?
Researchers, air monitoring professionals, environmental consultants, and industrial hygienists use the ZR-7250 AQMS to set up national and urban air monitoring networks, conduct environmental impact assessments, and to ensure sensitive receptors in the community are not at risk from air pollution.
What can ZR-7250 measure?
> Particulate Matter: PM10, PM2.5, PM1
> Gases: SO2, NOX, CO, O3
> Environmental: Temperature, humidity, noise, barometric pressure, wind speed and direction
Practical applications for the ZR-7250 AQMS include:
> Urban air monitoring networks
> National air monitoring networks
> Roadside air monitoring
> Industrial perimeter monitoring
> Environmental impact assessments
> Research and consultancy projects
> Short term hot spot monitoring
Features
> Continuous, simultaneous measurement of up to 10 common air pollutants and environmental parameters in real-time.
> The AQMS Series can be customized. A unique modular design increases flexibility and makes maintenance and servicing easy.
> The station can also be equipped with integrated calibration.
> Data is traceable back to international standards – USEPA (40 CFR Part 53) and EU (2008/50/EC).
> Remote data transmission, powerful data storage function up to one year.
Deliver Goods
Parameter |
CO |
SO2 |
NOx |
O3 |
Principle |
NDIR |
UV Fluorescence |
CLIA |
UV Spectrophotometry |
Range |
(0~50)umol/mol |
(0~500)umol/mol |
(0~500)nmol/mol |
(0~500)nmol/mol |
Sampling flowrate |
(800-1500)mL/min |
(500-1000)mL/min |
(450±45)mL/min |
800 mL/min |
The lowest detect limit |
≤0.5 umol/mol |
≤2 umol/mol |
≤0.5 nmol/mol |
≤1 nmol/mol |
Error |
±2%F.S. |
±5%F.S. |
±3%F.S. |
±2%F.S. |
Response |
≤4min |
≤5min |
≤120s |
≤30s |
Data storage |
250000 groups |
|||
Size |
(L494*W660*H188)mm |
|||
Weight |
15kg |
|||
Power supply |
AC(220±22)V,(50±1)Hz |
|||
Consumption |
≤300W |
≤300W |
≤700W |
≤300W
|
Parameter |
PM10/PM2.5/PM1 |
Principle |
Beta Attenuation method |
Range |
(0~1000)μg/m3 or(0~10000)μg /m3 |
Sampling flowrate |
16.7L/min |
Sampling cycle |
60min |
Atmospheric pressure |
(60~130)kPa |
Humidity |
(0~100)%RH |
Data storage |
365 day hourly concentration data |
Size |
(L324*W227*H390)mm |
Weight |
11kg(Sampling head included) |
Consumption |
≤150W |
Power supply |
AC(220±22)V,(50±1)Hz |