Microbial Limit Tester ZR-1140
product detail
Microbial Limit Tester (MLT) is used to determine whether a non-sterile pharmaceutical product complies with an established specification for microbial quality.
This is an important series of tests designed to limit the likelihood of any adverse incident occurring as a result of a high level of microorganisms being present in a non-sterile product. By setting limits on the number of bacteria present in the product under test you inherently reduce the likelihood of this occurring.
Application >>
Accessories
A complete Microbial Limit Tester consists of manifold, pump, funnels and membrane filters.
1. Host, Power line 2. Flame gun 3. Filter heads: Φ47mm, 4 pcs 4.Filter Bottle: Φ47mm, 10 pcs
5. Filter membrane:Φ47mm 0.45μm, 10 pcs
6.Manual 7.Qualification Certificate.
Features
> Built-in excellent diaphragm pump, low noise and reliable.
> Filtration membrane φ47mm (optional φ50mm, φ60mm).
> lSIP (Sanitize In Place) function.
> The 4 pump heads can be set timing independently. The pump heads can be dismantled easily for steam sterilization at 121℃.
> Built-in 4 high-performance diaphragm pumps, one pump one way, the direct drainage does not need a filter bottle.
> LCD touch screen, large data storage. The data can be exported to USB and printed by Bluetooth printer(optional).
Technical Parameter
Parameter |
Range |
Filter heads |
4 (1, 3, 6 optional) |
Membrane Diameter |
Φ47mm (Φ50mm, Φ60mm optional) |
Filter Bottle |
100ml |
Pore size of membrane |
0.45μm |
Dimention |
(415*250*130)mm |
Weight |
7kg |
Flowrate |
>1500ml/min(Empty) >600ml/min(Full load) |
Noise |
<65dB(A) |