I-ZR-1000A Mask Viral Filtration Efficiency (VFE) Detector
Ukucaciswa okuyinhloko komhloli we-ZR-1000A mask virus filtration performance (VFE) kuhambisana ne-YY/T1497-2016 , Indlela yegesi ephindwe kabili ukusampula kokuqhathanisa ngesikhathi esisodwa kuthuthukisa ukunemba kwesampula. Isebenza ekuhlolweni kokusebenza kokuhlunga kwegciwane ukusebenza kahle komnyango wokuqinisekisa i-metrological, izikhungo zocwaningo lwesayensi, abakhiqizi bemaski neminye iminyango efanele.
Izici
> Uhlelo lokuhlola ingcindezi engemihle luqinisekisa ukuphepha kwabaqhubi;
> Ikhabhinethi yengcindezi engemihle ifakwe iphampu ye-peristaltic, i-A kanye ne-B yezindlela ezimbili zesiteji sesithupha sika-Anderson sampler;
> Ukugeleza kwepompo ye-peristaltic kungasethwa;
> Ivolumu ye-nebulization ye-microbial aerosol generator ekhethekile ingasethwa futhi nokusebenza kwe-nebulization kuhle;
> Ilawulwa yi-microcomputer yezimboni enesivinini esikhulu;
> isikrini sokuthinta esinombala esingu-10.4-intshi esiphezulu;
> Imbobo ye-USB, esekela ukudluliswa kwedatha kwi-USB flash disk;
> Izibani ezikhanyayo ezikhangayo ekhabetheni;
> Iwundlu le-UV elakhelwe ngaphakathi ukuze libulawe amagciwane kanye nokuvalwa inzalo;
> Iswishi eyakhelwe ngaphakathi yokuvikela ukuvuza ukuze kuvikelwe ukuphepha kwabaqhubi;
> Ungqimba lwangaphakathi lwekhabhinethi lwenziwe ngensimbi engagqwali ngokucutshungulwa nokubunjwa jikelele, ungqimba lwangaphandle luyishidi eligoqwe elibandayo elifuthwe ngepulasitiki;
> Ukonga ukushisa kanye nokunciphisa ilangabi kusethwe phakathi kwezingqimba zangaphakathi nezingaphandle;Ingaphambili Umnyango wengilazi ovulelekayo wangaphambili ulungele umqhubi ohlolayo ukuba awubheke futhi awusebenzise;
> Abakaki abasusekayo abanobude obulungisekayo;
> I-caster ingasetshenziselwa ukusekela nokunyakaza.
Diliva Izimpahla
Ipharamitha Eyinhloko | Ipharamitha Range | Isixazululo | I-MPE (Iphutha Eliphezulu Elivumelekile) |
Ukugeleza kwesampula yesiteshi A | (5-35) L/min | 0.1L/min | ±2.5% |
Ukugeleza kwesampula yesiteshi B | (5-35) L/min | 0.1L/min | ±2.5% |
Izinga lokugeleza kwe-Nebulization | (6-8) L/min | 0.1L/min | ±5.0% |
Ukugeleza kwepompo ye-Peristaltic | (0.006-3.0) mL/min | 0.001ml/min | ±2.5% |
Ingcindezi yangaphambili ye-A-path flowmeter | (-50~0) kPa | 0.01kPa | ±2.5% |
Ingcindezi yangaphambili ye-B-path flowmeter | (-50-0) kPa | 0.01kPa | ±2.5% |
Ingcindezi yangaphambili ye-spray flowmeter | (0-300) kPa | 0.1kPa | ±2.5% |
Ingcindezi engemihle egumbini le-aerosol | (-90 ~ 120) Kuhle | 0.1Pa | ±2.0% |
izinga lokushisa lokusebenza | (0-50) ºC | ||
Ingcindezi engemihle yeKhabhinethi | (-50 ~ 200) Kuhle | ||
Umthamo wokugcina idatha | > Iqembu le-100000 | ||
Izici zesihlungi se-HEPA | Ukuphumelela kokuhlunga kwezinhlayiya ezingaphezu kuka-0.3μm ≥99.99% | ||
Isilinganiso esikhulu se-aerosol generator | Isilinganiso sobubanzi bezinhlayiyana (3.0 ±0.3) μm, ukuchezuka okujwayelekile kwejiyomethri ≤ 1.5 | ||
Ubukhulu bekamelo le-aerosol | (ubude 600× ububanzi 85× ukujiya 3) mm | ||
Ukugeleza kwe-Ventilation OF i-negative pressure cabinet | ≥5m3/min | ||
Ubukhulu bomnyango wekhabethe lengcindezi engalungile | (ubude 1000× ububanzi 730)mm | ||
Usayizi womsingathi | (ubude 1180×ububanzi 650×ubude 1300)mm | ||
Usayizi wabakaki | (ubude 1180× ububanzi 650× ubude 600)mm, ukuphakama adjustable ngaphakathi 100mm | ||
Ukunikezwa kwamandla okusebenzayo | I-AC220V±10%,50Hz | ||
Umsindo wezinsimbi | |||
Isisindo somshini wonke | Cishe 250kg | ||
Ukusetshenziswa kwamandla sekukonke |