I-ZR-1000C Mask Bacterial Filtration Efficiency (BFE) tester

Inkcazelo emfutshane:

Iimpawu eziphambili zeModel ZR-1000 imaski ibhaktheriya yokuhluza ukusebenza kakuhle komvavanyi (BFE) ayifezekisi kuphela iimfuno ze-B.1.1 kwi-YY0469-2011 , kodwa ukwahlangabezana neemfuno ze-ASTMF2100, ASTMF2101 kunye nemigangatho ye-EN14683 yaseYurophu.


  • Isampulu yoMlinganiselo wokuHamba ka-A:28.3L/min
  • Isampulu yoMlinganiselo wokuHamba ka-B:28.3L/min
  • Isantya sokuhamba kweNebulization:(0.1~10)L/min
  • Ukuhamba kwempompo yePeristaltic:(0.001~3.0) mL/min
  • Lilonke inani lesampulu yolawulo lomgangatho olungileyo:(2200±500) CFU/mL
  • Ubungakanani bomamkeli:(ubude 1050× ububanzi 635× ukuphakama 1310) mm
  • Ubunzima bomatshini wonke:Malunga ne-180 kg
  • Iinkcukacha zeMveliso

    Inkcazo

    Iimpawu eziphambili zeModel ZR-1000 imaski ibhaktheriya yokuhluza ukusebenza kakuhle komvavanyi (BFE) ayifezekisi kuphela iimfuno ze-B.1.1 kwi-YY0469-2011 , kodwa ukwahlangabezana neemfuno ze-ASTMF2100, ASTMF2101 kunye nemigangatho ye-EN14683 yaseYurophu. Ngaphezu koko, senze uphuculo olutsha olusekwe koku, kunye nesampulu ephindwe kabini yokuthelekisa igesi ngaxeshanye, siphucule ukuchaneka kwesampulu. Isebenza kuvavanyo lwentsebenzo yobuchule bokuhluza ibhaktiriya yemaski lisebe lokuqinisekisa i-metrological, amaziko ophando lwenzululwazi, abavelisi bemaski kunye namanye amasebe afanelekileyo.

    Iimbonakalo

    > Inkqubo yovavanyo loxinzelelo olubi iqinisekisa ukhuseleko lwabaqhubi;

    > Ikhabhinethi yoxinzelelo olubi ixhotyiswe ngepompo ye-peristaltic, i-A kunye ne-B yeendlela ezimbini zenqanaba lesithandathu i-Anderson sampler;

    > Ukuhamba kwempompo ye-peristaltic kunokusetwa;

    > Isantya sokuhamba kwejenereyitha ye-aerosol ekhethekileyo ye-microbial inokumiselwa.

    > Ukusebenza kakuhle kwe-nebulization, ubukhulu be-aerosol buhambelana nemigangatho yehlabathi;

    > Ukulawulwa yi-microcomputer yemizi-mveliso enesantya esiphezulu;

    > I-10.4-intshi ye-intshi ye-intshi ephezulu ekhanyayo enemibala yesikrini sokuchukumisa;

    > Izibuko le-USB, inkxaso yokudluliselwa kwedatha kwi-USB flash disk;

    > Inkxaso yoshicilelo lweBluetooth;

    Izibane ezakhelwe ngaphakathi ezikhanyayo;

    > Iswitshi yokukhusela ukuvuza ukukhusela ukhuseleko lwabaqhubi;

    > Uluhlu lwangaphakathi lwekhabhinethi lwenziwe ngensimbi engenasici, kwaye uluhlu lwangaphandle luyi-cold roll sheet sheeted by plastic;

    > Ugcino lobushushu kunye nokubuyiswa komlilo phakathi kwamaleko angaphakathi nangaphandle;

    > Ucango lwangaphambili lweglasi oluvulelekayo lulungele ukuba umsebenzisi wovavanyo aluqwalasele kwaye alusebenzise;

    > Isibambi esinokususwa, kunye ne-caster ingasetyenziselwa ukuxhasa kunye nokuhamba.

    > Isikrini esijikelezayo esinokuhlehliswa, esilungele ukusebenza;

    > Ikhabhinethi yoxinzelelo olubi inebhodi engundoqo yokusebenza eyakhelweyo, ekwazi ukumelana ne-corrosion kunye ne-UV-resistant, yenza kube lula ukusebenza.

    Imigangatho

    EN14683-2019

    I-ASTM F2100, STM F2101

    YY 0469-2011

    YY/T 0969-2013

    GB/T 38880-2020

    T/CNTAC 55—2020 T/CNITA 09104——2020

    T/CTCA 7-2019 

    Hambisa iMpahla

    ukuhambisa iimpahla ElamaTaliyane
  • Ngaphambili:
  • Okulandelayo:

  • Iiparamitha eziphambili

    Uluhlu lweParameter

    Isigqibo

    I-MPE (Eyona mpazamo iphezulu ivumelekileyo)

    Isampulu yoMlinganiselo wokuHamba ka-A

    28.3L/min

    0.1L/min

    ±2.0%

    Isampulu yoMgangatho wokuHamba ka-B

    28.3L/min

    0.1L/min

    ±2.0%

    Isantya sokuhamba kweNebulization

    (0.1~10)L/min

    0.1L/min

    ±5.0%

    Ukuhamba kwepompo yePeristaltic

    (0.001~3.0) mL/min

    0.01mL/min

    ±2.0%

    Uxinzelelo lwangaphambili lwe-A-path flowmeter

    (-20~0)kPa

    0.1kPa

    ±2.5%

    Uxinzelelo lwangaphambili lwe-B-path flowmeter

    (-20~0)kPa

    0.01kPa

    ±2.5%

    Uxinzelelo lwangaphambili lwe-nebulizer

    (0~50)kPa

    0.01kPa

    ±2.5%

    Uxinzelelo olubi kwigumbi le-aerosol

    (-90~-120) Kulungile

    0.1Pa

    ±1.0%

    Ubushushu bokusebenza

    (0~50)℃

    Uxinzelelo olubi lweKhabhinethi

    (-50~-200) Kulungile

    Ukukwazi ukugcina idatha

    >500000 amaqela

    Iimpawu zesihluzo se-HEPA

    ≥ 99.995%@0.3μm, ≥ 99.995%@0.12μm

    Ubunzima obuphakathi kwe-aerosol

    I-avareji yamasuntswana idiameter ngu (3.0 ± 0.3)μm, ukutenxa umgangatho wejometri ≤ 1.5

    Ubukhulu besuntswana elithathiweyo leendlela ezimbini ze-6-inqanaba le-Anderson sampler

    Inqanaba lokuqala > 7 μm, isigaba se-2 (4.7-7) μm, isigaba se-3 (3.3-4.7) μm, isigaba se-4 (2.1-3.3) μm, isigaba se-5 (1.1-2.1) μm, isigaba se-6 (0.6-1.1) μm

    Ubungakanani begumbi le-aerosol

    (ubude 600× ububanzi 85× ubukhulu 3) mm

    Lilonke inani lesampulu yolawulo olulungileyo

    (2200±500) CFU/mL

    Isantya sokuhamba komoya wekhabhinethi yoxinzelelo olubi

    ≥5m3/min

    Imilinganiselo yocango lwekhabhathi yoxinzelelo olubi

    (ubude 1100× ububanzi 780) mm

    Ubungakanani bomamkeli

    (ubude 1050× ububanzi 635× ukuphakama 1310) mm

    Ubungakanani besibiyeli

    (ubude 1050× ububanzi 635× ukuphakama 650) mm

    Ukunikezwa Amandla

    AC220V±10%,50Hz

    Ingxolo yesixhobo

    <65dB(A)

    Ubunzima bomatshini wonke

    Malunga ne-180 kg

    Ukusetyenziswa kwamandla ngokubanzi

    <500W

    Bhala umyalezo wakho apha kwaye uwuthumele kuthi