Manufacturing Companies for Kn95 Effect Tester - Mask Bacterial Filtration Efficiency (BFE) Tester – Junray

    Short Description:


    Product Detail

    Related Video

    Feedback (2)

    With a positive and progressive attitude to customer's interest, our enterprise continually improves our products excellent to meet the wants of customers and further focuses on safety, reliability, environmental requirements, and innovation of Kn95 Protection Effect Test Machine , Microbial Aerosol Sampling , Mask Air Resistance Test , The continual availability of significant grade merchandise in combination with our excellent pre- and after-sales support ensures strong competitiveness in an increasingly globalized market.
    Manufacturing Companies for Kn95 Effect Tester - Mask Bacterial Filtration Efficiency (BFE) Tester – Junray Detail:

    price:US $48,000-90,000/ Piece

    Introduction

    Key specifications of Model ZR-1000 mask bacteria filtration efficiency tester (BFE) not only meet the requirements of B.1.1 in YY0469-2011 <technical requirements for medical surgical masks>, but also meet the requirements of ASTMF2100, ASTMF2101 and European EN14683 standards. What’s more, we made innovative improvements based on this, with double gas contrast sampling method at the same time, improve the precision of sampling. It is applicable to the performance test of the mask bacterial filtration efficiency by the metrological verification department, scientific research institutes, mask manufacturers and other relevant departments.

    Features

    The negative pressure test system ensures the safety of operators;

    The negative pressure cabinet is equipped with peristaltic pump, A and B two-path six stage Anderson sampler;

    The flow of peristaltic pump can be set;

    The flow rate of microbial aerosol generator can be set.

    Good nebulization performance;

    Controlled by embedded high-speed industrial microcomputer;

    10.4-inch industrial high brightness colored touch screen;

    USB port, supporting data transfer of USB flash disk;

    Built-in high brightness lights;

    Built in leakage protection switch to protect the safety of operators;

    The inner layer of the cabinet is made of stainless steel, and the outer layer is cold-rolled sheet sprayed by plastic;

    Heat preservation and flame retardant is set between inner and outer layers;

    The front openable glass door is convenient for the experimental operator to observe and operate;

    Detachable bracket with adjustable height;

    The caster can be used for supporting and moving.

    Standards

    EN14683-2019

    ASTM F2100, STM F2101

    YY 0469-2011

    YY/T 0969-2013

    GB/T 38880-2020

    T/CNTAC 55—2020 T/CNITA 09104——2020

    T/CTCA 7—2019

    Specifications

    Main parameters

     Parameter range 

    Resolution

    MPE (Maximum Permissible Error)

    Sampling Flow Rate of A 28.3L/min 0.1L/min ±2.5%
    Sampling Flow Rate of B 28.3L/min 0.1L/min ±2.5%
    Nebulization flow rate (8~10)L/min 0.1L/min ±5.0%
    Peristaltic pump flow

    (0.006~3.0) mL/min

    0.001mL/min ±2.5%
    Front pressure of A-path flowmeter (-20~0)kPa 0.1kPa ±2.5%
    Front pressure of B-path flowmeter (-20~0)kPa 0.1kPa ±2.5%
    Front pressure of nebulizer (0~300)kPa 0.1kPa ±2.5%
    Negative pressure in aerosol chamber (-90~-120)Pa 1Pa ±2.0%
    Operation temperature (0~50)℃
    Cabinet negative pressure (-50~-200)Pa
    Data storage capability >100000 groups
    Characteristics of HEPS filter Filtration efficiency for particles is above 0.3μm ≥ 99.99%
    Mean mass diameter of aerosols Average particle diameter is (3.0 ± 0.3) μm, geometric standard deviation ≤ 1.5
    Captured particle size of Two-path 6-stage Anderson sampler 1st stage > 7 μm, 2nd stage (4.7-7) μm,3rd stage (3.3-4.7) μm, 4th stage(2.1-3.3) μm, 5th stage (1.1-2.1) μm, 6th stage (0.6-1.1) μm
    Size of aerosol chamber (length 600× diameter 85× thickness 3) mm
    Total number of positive quality control sampler (2200±500) CFU
    Ventilation flow rate of negative pressure cabinet ≥5m3/min
    Dimensions of negative pressure cabinet door (length 1000× width 730) mm
    Host size (length 1180× width 650× height 1300) mm
    Bracket size (length 1180× width 650× height 600) mm, height adjustable within 100mm
    Power supply AC220V±10%,50Hz
    Instrument noise <65dB(A)
    Weight of whole machine About 250 kg
    Overall power consumption <1500W

    Deliver Goods

    deliver goodsItaly
    Product detail pictures:


    Related Product Guide:

    "Based on domestic market and expand overseas business" is our development strategy for Manufacturing Companies for Kn95 Effect Tester - Mask Bacterial Filtration Efficiency (BFE) Tester – Junray , The product will supply to all over the world, such as: Kuwait, Dubai, Kuwait, Facing fierce global market competition, we have launched the brand building strategy and updated the spirit of "human-oriented and faithful service", with an aim to gain global recognition and sustainable development.
    The factory has advanced equipment, experienced staffs and good management level, so product quality had assurance, this cooperation is very relaxed and happy!
    5 StarsBy Elsie from Birmingham - 2018.12.30 10:21
    The factory has advanced equipment, experienced staffs and good management level, so product quality had assurance, this cooperation is very relaxed and happy!
    5 StarsBy Kristin from Qatar - 2018.06.03 10:17
    Write your message here and send it to us