Manufacturer of Medical Mask Bfe Test Machine - Mask Bacterial Filtration Efficiency (BFE) Tester – Junray
Manufacturer of Medical Mask Bfe Test Machine - Mask Bacterial Filtration Efficiency (BFE) Tester – Junray Detail:
price:US $48,000-90,000/ Piece
Introduction
Key specifications of Model ZR-1000 mask bacteria filtration efficiency tester (BFE) not only meet the requirements of B.1.1 in YY0469-2011 <technical requirements for medical surgical masks>, but also meet the requirements of ASTMF2100, ASTMF2101 and European EN14683 standards. What’s more, we made innovative improvements based on this, with double gas contrast sampling method at the same time, improve the precision of sampling. It is applicable to the performance test of the mask bacterial filtration efficiency by the metrological verification department, scientific research institutes, mask manufacturers and other relevant departments.
Features
The negative pressure test system ensures the safety of operators;
The negative pressure cabinet is equipped with peristaltic pump, A and B two-path six stage Anderson sampler;
The flow of peristaltic pump can be set;
The flow rate of microbial aerosol generator can be set.
Good nebulization performance;
Controlled by embedded high-speed industrial microcomputer;
10.4-inch industrial high brightness colored touch screen;
USB port, supporting data transfer of USB flash disk;
Built-in high brightness lights;
Built in leakage protection switch to protect the safety of operators;
The inner layer of the cabinet is made of stainless steel, and the outer layer is cold-rolled sheet sprayed by plastic;
Heat preservation and flame retardant is set between inner and outer layers;
The front openable glass door is convenient for the experimental operator to observe and operate;
Detachable bracket with adjustable height;
The caster can be used for supporting and moving.
Standards
EN14683-2019
ASTM F2100, STM F2101
YY 0469-2011
YY/T 0969-2013
GB/T 38880-2020
T/CNTAC 55—2020 T/CNITA 09104——2020
T/CTCA 7—2019
Specifications
Main parameters |
Parameter range |
Resolution |
MPE (Maximum Permissible Error) |
Sampling Flow Rate of A | 28.3L/min | 0.1L/min | ±2.5% |
Sampling Flow Rate of B | 28.3L/min | 0.1L/min | ±2.5% |
Nebulization flow rate | (8~10)L/min | 0.1L/min | ±5.0% |
Peristaltic pump flow |
(0.006~3.0) mL/min |
0.001mL/min | ±2.5% |
Front pressure of A-path flowmeter | (-20~0)kPa | 0.1kPa | ±2.5% |
Front pressure of B-path flowmeter | (-20~0)kPa | 0.1kPa | ±2.5% |
Front pressure of nebulizer | (0~300)kPa | 0.1kPa | ±2.5% |
Negative pressure in aerosol chamber | (-90~-120)Pa | 1Pa | ±2.0% |
Operation temperature | (0~50)℃ | ||
Cabinet negative pressure | (-50~-200)Pa | ||
Data storage capability | >100000 groups | ||
Characteristics of HEPS filter | Filtration efficiency for particles is above 0.3μm ≥ 99.99% | ||
Mean mass diameter of aerosols | Average particle diameter is (3.0 ± 0.3) μm, geometric standard deviation ≤ 1.5 | ||
Captured particle size of Two-path 6-stage Anderson sampler | 1st stage > 7 μm, 2nd stage (4.7-7) μm,3rd stage (3.3-4.7) μm, 4th stage(2.1-3.3) μm, 5th stage (1.1-2.1) μm, 6th stage (0.6-1.1) μm | ||
Size of aerosol chamber | (length 600× diameter 85× thickness 3) mm | ||
Total number of positive quality control sampler | (2200±500) CFU | ||
Ventilation flow rate of negative pressure cabinet | ≥5m3/min | ||
Dimensions of negative pressure cabinet door | (length 1000× width 730) mm | ||
Host size | (length 1180× width 650× height 1300) mm | ||
Bracket size | (length 1180× width 650× height 600) mm, height adjustable within 100mm | ||
Power supply | AC220V±10%,50Hz | ||
Instrument noise | <65dB(A) | ||
Weight of whole machine | About 250 kg | ||
Overall power consumption | <1500W |
Deliver Goods


Product detail pictures:



Related Product Guide:
Well-run devices, expert profits group, and better after-sales companies; We have been also a unified huge family, everybody keep on with the organization worth "unification, determination, tolerance" for Manufacturer of Medical Mask Bfe Test Machine - Mask Bacterial Filtration Efficiency (BFE) Tester – Junray , The product will supply to all over the world, such as: Lebanon, Salt Lake City, Slovakia, Our monthly output is more than 5000pcs. We have set up a strict quality control system. Please feel free to contact us for further information. We hope that we can establish long-term business relationships with you and carry out business on a mutually beneficial basis. We are and will be always trying our best to serve you.

Staff is skilled, well-equipped, process is specification, products meet the requirements and delivery is guaranteed, a best partner!