Manufactur standard Mask Tightness Rating - Mask Bacterial Filtration Efficiency (BFE) Tester – Junray

Short Description:


Product Detail

Related Video

Feedback (2)

Our eternal pursuits are the attitude of "regard the market, regard the custom, regard the science" as well as the theory of "quality the basic, have confidence in the very first and management the advanced" for Animal Aerosol Exposure Device , Pfe , Face Mask Filtration Tester , We just not only deliver the high-quality to our customers, but far more even important is our greatest service along with the competitive price tag.
Manufactur standard Mask Tightness Rating - Mask Bacterial Filtration Efficiency (BFE) Tester – Junray Detail:

price:US $48,000-90,000/ Piece

Introduction

Key specifications of Model ZR-1000 mask bacteria filtration efficiency tester (BFE) not only meet the requirements of B.1.1 in YY0469-2011 <technical requirements for medical surgical masks>, but also meet the requirements of ASTMF2100, ASTMF2101 and European EN14683 standards. What’s more, we made innovative improvements based on this, with double gas contrast sampling method at the same time, improve the precision of sampling. It is applicable to the performance test of the mask bacterial filtration efficiency by the metrological verification department, scientific research institutes, mask manufacturers and other relevant departments.

Features

The negative pressure test system ensures the safety of operators;

The negative pressure cabinet is equipped with peristaltic pump, A and B two-path six stage Anderson sampler;

The flow of peristaltic pump can be set;

The flow rate of microbial aerosol generator can be set.

Good nebulization performance;

Controlled by embedded high-speed industrial microcomputer;

10.4-inch industrial high brightness colored touch screen;

USB port, supporting data transfer of USB flash disk;

Built-in high brightness lights;

Built in leakage protection switch to protect the safety of operators;

The inner layer of the cabinet is made of stainless steel, and the outer layer is cold-rolled sheet sprayed by plastic;

Heat preservation and flame retardant is set between inner and outer layers;

The front openable glass door is convenient for the experimental operator to observe and operate;

Detachable bracket with adjustable height;

The caster can be used for supporting and moving.

Standards

EN14683-2019

ASTM F2100, STM F2101

YY 0469-2011

YY/T 0969-2013

GB/T 38880-2020

T/CNTAC 55—2020 T/CNITA 09104——2020

T/CTCA 7—2019

Specifications

Main parameters

 Parameter range 

Resolution

MPE (Maximum Permissible Error)

Sampling Flow Rate of A 28.3L/min 0.1L/min ±2.5%
Sampling Flow Rate of B 28.3L/min 0.1L/min ±2.5%
Nebulization flow rate (8~10)L/min 0.1L/min ±5.0%
Peristaltic pump flow

(0.006~3.0) mL/min

0.001mL/min ±2.5%
Front pressure of A-path flowmeter (-20~0)kPa 0.1kPa ±2.5%
Front pressure of B-path flowmeter (-20~0)kPa 0.1kPa ±2.5%
Front pressure of nebulizer (0~300)kPa 0.1kPa ±2.5%
Negative pressure in aerosol chamber (-90~-120)Pa 1Pa ±2.0%
Operation temperature (0~50)℃
Cabinet negative pressure (-50~-200)Pa
Data storage capability >100000 groups
Characteristics of HEPS filter Filtration efficiency for particles is above 0.3μm ≥ 99.99%
Mean mass diameter of aerosols Average particle diameter is (3.0 ± 0.3) μm, geometric standard deviation ≤ 1.5
Captured particle size of Two-path 6-stage Anderson sampler 1st stage > 7 μm, 2nd stage (4.7-7) μm,3rd stage (3.3-4.7) μm, 4th stage(2.1-3.3) μm, 5th stage (1.1-2.1) μm, 6th stage (0.6-1.1) μm
Size of aerosol chamber (length 600× diameter 85× thickness 3) mm
Total number of positive quality control sampler (2200±500) CFU
Ventilation flow rate of negative pressure cabinet ≥5m3/min
Dimensions of negative pressure cabinet door (length 1000× width 730) mm
Host size (length 1180× width 650× height 1300) mm
Bracket size (length 1180× width 650× height 600) mm, height adjustable within 100mm
Power supply AC220V±10%,50Hz
Instrument noise <65dB(A)
Weight of whole machine About 250 kg
Overall power consumption <1500W

Deliver Goods

deliver goodsItaly
Product detail pictures:

Manufactur standard Mask Tightness Rating - Mask Bacterial Filtration Efficiency (BFE) Tester – Junray detail pictures

Manufactur standard Mask Tightness Rating - Mask Bacterial Filtration Efficiency (BFE) Tester – Junray detail pictures

Manufactur standard Mask Tightness Rating - Mask Bacterial Filtration Efficiency (BFE) Tester – Junray detail pictures


Related Product Guide:

Our merchandise are broadly identified and trusted by end users and can satisfy continually developing economic and social requires for Manufactur standard Mask Tightness Rating - Mask Bacterial Filtration Efficiency (BFE) Tester – Junray , The product will supply to all over the world, such as: British, South Africa, Denver, Our qualified engineering team will usually be prepared to serve you for consultation and feedback. We've been able to also deliver you with absolutely free samples to meet your needs. Best efforts might be made to offer you the ideal service and items. For anyone who is interested in our company and products, be sure to make contact with us by sending us emails or contact us right away. In order to know our solutions and organization. ar more, you can come to our factory to determine it. We are about to usually welcome guests from around the globe to our corporation. o create small business relations with us. Please genuinely feel no cost to speak to us for enterprise. nd we believe we have been gonna share the most effective trading practical experience with all our merchants.
This company conforms to the market requirement and joins in the market competition by its high quality product, this is an enterprise that have Chinese spirit.
5 StarsBy Elaine from Myanmar - 2018.06.09 12:42
Speaking of this cooperation with the Chinese manufacturer, I just want to say"well dodne", we are very satisfied.
5 StarsBy Natalie from Milan - 2018.06.28 19:27
Write your message here and send it to us