Good quality Six-Stage Microorganism Sampling - ZR-2050A Airborne microbe sampler – Junray
Good quality Six-Stage Microorganism Sampling - ZR-2050A Airborne microbe sampler – Junray Detail:
US $30,00-45,00 / Piece
Introduction
ZR-2050A airborne microbe sampler is a high efficiency single stage multiple aperture impact sampler, this instrument bases on Anderson impacting principle, impacting speed is 10.8 m/s, it could capture all particles larger than 1 1μm. This instrument will draw the air through multiple aperture sampling head, impact to Φ90mm petri dish, the airborne microbes will be captured to agar medium. This instrument is widely applied in pharmaceutical industry, food industry, drug inspection institute, centers of disease control, health and epidemic prevention, hospitals and other related industries and departments
Standards
GMP GOOD MANUFACTURING PRACTICES
ISO 14698-1/2 Cleanrooms and associated controlled environments — Biocontamination control
GB/T 16293-2010 Test method for airborne microbe in cleanroom (zone) of the pharmaceutical industry
Features
Original imported fan, electronic flow sensor, high flow control precision.
Automatically stop sampling when can’t reach sampling flow(Low power of the sampling head is blocked) to guarantee the sampling precision.
Build-in large capacity lithium battery, keep sampling up to 6 hours.
Query historical data anytime and support export to USB flash disk.
Dot matrix LCD display, Chinese and English bilingual menu.
Realtime display clock.
Power saving function, automatically adjust backlight brightness and automatically shut down when long time no operation.
Anodized aluminum shell, solid and durable, stylish and beautiful;
Sampling head could be adjusted on horizonal and vertical direction.
Special tripod, the sampling height is adjustable.
Sampling head for compress air (Optional).
Specification
| The main parameter | Parameter range | Resolution | Maximum permitted error (MPE) |
| Sampling flow | 100L/min | 1L/min | ±2.5% |
| Working temperature | (-20~50)℃ | ||
| Range of Sampling volume | (20~5000)L | ||
| Specification of petri dish | Ф90mm | ||
| Instrument noise | <60dB(A) | ||
| Battery working time | >6h | ||
| Power adaptor | Input AC100~240V 50/60Hz Output DC15V 3A | ||
| Size of host | (length 130×width 110×height210)mm | ||
| Weight of host | about 1.5kg | ||
| Host power consumption | <15W | ||
Deliver Goods


Product detail pictures:
Related Product Guide:
We've been proud with the significant shopper fulfillment and wide acceptance due to our persistent pursuit of top of the range both of those on solution and repair for Good quality Six-Stage Microorganism Sampling - ZR-2050A Airborne microbe sampler – Junray , The product will supply to all over the world, such as: Comoros, United States, America, "Make the women more attractive "is our sales philosophy. "Being customers' trusted and preferred brand supplier" is the goal of our company. We are strict with every part of our work. We sincerely welcome friends to negotiate business and start cooperation. We hope to join hands with friends in different industries to create a brilliant future.