Good quality Six-Stage Microorganism Sampling - ZR-1013 Biosafety cabinet quality tester – Junray Detail:
US $40,000-60,000 / Piece
Introduction
ZR-1013 Biosafety cabinet quality tester adopts potassium iodide (KI) method to test the protection performance of biosafety cabinet class II, conforms to JJF 1815-2020 <Calibration Specification for Class II biosafety Cabinets> and related standards. Build in background test, operator protection, product protection and cross protection four working mode. It is meanly used to test whether the aerosol leaked to outside of the environment, whether the outer agents could enter the work area, whether the contamination among samples is minimized. It is applicable to the protection performance test of class II biosafety cabinets by metrological verification departments, scientific research institutes and biosafety cabinet manufacturers.
Features
8-inch colored high definition touch screen, the contents are intuitive and operation is easy.
Equipped special aerosol sampler, could capture special size aerosol.
Equipped calibrated special KI aerosol generator, the rotation speed is stable.
Four path independent high precision sampling module, automatic flow control, don’t need to adjust pressure to stabilize flow.
Test results could be exported to USB drive or printed by Bluetooth printer.
High precision liquid supply module, guarantee a stable liquid flow.
The automatic aerosol control port adopts PID arithmetic, control by real time feedback generator rotation speed.
Operator protection, product protection and cross protection test three test modes could start after just press the button.
Standards
YY 0569-2011 Class II biological safety cabinets
JJF 1815-2020 Calibration specification for of class II biosafety cabinet
DB52T 1254-2017 Technical practice for testing biosafety cabinets
Specification
|
The main parameters |
Parameter range |
Resolution |
Maximum permitted error (MPE) |
|
Sampling flow |
100L/min |
0.1L/min |
±2.0% |
|
Rotation speed of aerosol generator |
28000r/min |
/ |
±500r/min |
|
Diameter of rotating plate on aerosol generator |
38mm |
/ |
/ |
|
Highest position of X1、Y1 |
1000mm |
||
|
Sampling membrane |
Diameter 25mm,Aperture 3μm |
||
|
Disturbing cylinder |
Diameter 63mm,Length 1100mm |
||
|
Noise |
<65dB(A) |
||
|
Working temperature |
0-40℃ |
||
|
Power supply |
AC220V±10%,50Hz |
||
|
Overall size |
(length 450×width 380×height 720)mm |
||
|
Overall power |
< 400W |
||
|
Overall weight |
About 23kg |
||
Deliver Goods


Product detail pictures:
Related Product Guide:
We have state-of-the-art tools. Our products are exported towards the USA, the UK and so on, enjoying a fantastic reputation amongst customers for Good quality Six-Stage Microorganism Sampling - ZR-1013 Biosafety cabinet quality tester – Junray , The product will supply to all over the world, such as: Armenia, Albania, Kuwait, Our qualified engineering team will usually be prepared to serve you for consultation and feedback. We are able to also deliver you with absolutely free samples to meet your needs. Best efforts might be made to offer you the ideal service and products. For anyone who is interested in our company and items, please make contact with us by sending us emails or contact us right away. In order to know our solutions and organization. ar more, you can come to our factory to determine it. We are going to usually welcome guests from around the globe to our corporation. o create small business relations with us. Please really feel no cost to speak to us for enterprise. nd we believe we are going to share the most effective trading practical experience with all our merchants.