Introduction
ZR-1006 mask particulate filtration efficiency and air flow resistance tester is applicable to test the particle filtration efficiency and airflow resistance of masks and filter materials for centers of medical device inspection, security inspection center, drug inspection center, centers of diseases control and prevention, center of textile inspection, hospitals and mask R&D manufacturers.
Features
High definition LCD colored touch screen makes contents more intuitive and operation easier.
The equipped dedicated salt aerosol generators can generate aerosol in specific size and concentration.
Equipped multiple series of special fixtures are applicable to the test of various types of masks
Embedded long life photometer module, automatically count sampling time and prompt light path cleaning
To prevent human intervention, it can automatically calculate filtration efficiency and test airflow resistance of masks.
The embedded high-precision electronic flowmeter and high-performance sampling pump can insure the stability of flow.
Embedded compressor has the function of automatic pneumatic clamping.
Equipped with electrostatic removal device
The automatic infrared anti-pinch protection function to protect personnel safety.
No aerosol leakage, strong operator protection
Data of detection can be exported by USB disk or printed by Bluetooth printer.
Optional oil or salt aerosol generator.
Standards
GB/T 32610-2016 Technical specification of daily protective mask
GB/2626-2019 Respiratory protection——Non-powered air-purifying particle respirator
GB 19082-2009 Technical requirements for single-use protective clothing for medical use
GB 19083-2010 Technical requirements for protective face mask for medical use
TAJ 1001-2015 PM2.5 Protective mask
T/CNTAC 55—2020 T/CNITA 09104——2020 Civil sanitary mask
T/CTCA 7—2019 Ordinary protective mask
YY 0469-2011 Surgical mask
EN 149 Respiratory protective devices – Filtering half masks to protect against particles – Requirements, testing, marking
NIOSH 42 CFR Part 84 Respiratory Protective Devices
Specifications
| The Main Parameter | Parameter Range | Resolution | Maximum Permissible Error |
|
Median diameter of salt aerosol |
(0.075) μm |
/ |
±0.02 |
|
Median diameter of paraffin aerosol (Optional) |
(0.185) μm |
/ |
±0.02 |
|
Test flow |
(8~100) L/min |
0.1 L/min | ±2.0% |
|
Range of pressure detection |
(0~2500) Pa | 0.1 Pa | ±2.0% |
|
Range of concentration detection |
(0.001-200) mg/m3 | 0.001 mg/m3 | |
|
Precision of test |
/ | / | 1%FS |
|
Repeatability of test |
/ | / | 2% |
|
Operation temperature |
0~50℃ | ||
|
Instrument noise |
<60 dB(A) | ||
|
Power supply |
AC 220 V±10%,50 Hz | ||
|
Host size(L×W×H) |
(Length 550 × Width 550 × Height 1630) mm | ||
|
Power consumption |
<300 W | ||
|
Weight of host |
About 90 kg | ||
Deliver Goods


Product detail pictures:
Related Product Guide:
owing to fantastic assistance, a variety of high quality goods, aggressive rates and efficient delivery, we love a very good popularity among our customers. We are an energetic firm with wide market for China Manufacturer for Face Mask Test Instrument - ZR-1006 Mask Particulate Filtration Efficiency and Air flow Resistance Tester – Junray , The product will supply to all over the world, such as: Mecca, Belarus, Morocco, By adhering to the principle of "human oriented, winning by quality", our company sincerely welcomes merchants from at home and abroad to visit us, talk business with us and jointly create a brilliant future.
By Elaine from Australia - 2017.12.31 14:53
The quality of the products is very good, especially in the details, can be seen that the company work actively to satisfy customer's interest, a nice supplier.
By Gloria from Norwegian - 2018.06.18 19:26